| Display | Color LCD Display 5.7", 320X240 pixels |
| Pulse Rate |
Range : 30 ~ 300 bpm (SpO2) 30 ~ 300 bpm (EKG) Accuracy Heart Range : ± 3 bpm Alarm Range : 30 ~ 250 bpm (Full range adjustable) Default Setting: Upper Limit 250 bpm and Lower Limit 30 bpm |
| NIBP |
Method : Oscillometric Range : Neonate Mode 30 ~ 150 mmHg Adult Mode 30 ~ 250mmHg Accuracy : ± 5 mmHg Alarm Range: Systolic (90 ~ 200 mmHg) Diastolic (40 ~ 105 mmHg) Default Setting: Systolic: Upper Limit 180 mmHg and Lower Limit 80 mmHg Diastolic: Upper Limit 100 mmHg and Lower Limit 90 mmHg Cuff : Different size for adult , children and neonate Parameters : Systolic, Diastolic and MAP pressure |
| EKG |
One channel, 3-lead Configuration : 3-lead Lead Selection : Lead I Display Gain : 0.5 , 1 , 2 (mv/cm) Defibrillation Protection |
| EtCO2 (Optional) |
Method: Side Stream Technology Measurement : Non-dispersive IR Absorption Range : 0~100mmHg , STPD Accuracy EtCO2 : 0 ~ 38 mmHg ± 2 mmHg 39 ~ 100 mmHg < 10% of reading Alarm Range : 0 ~ 50 mmHg adjustable Default Setting: Upper Limit 50 mmHg and Lower Limit 0 mmHg Flow Rate : 75 , 150 , 250 ml/min (Default setting 75 ml/min) |
| Respiration (Optional) |
Range : 0 to 99 min-1 Accuracy : ± 1% Alarm Range : 0~99 min-1, adjustable Default Setting: Upper Limit 99 min-1 and Lower Limit 0 min-1 |
| Temperature (Optional) |
Temperature Range: 20°C ~ 45°C Resolution : 0.1°C Channel: One-Channel (extend to Two-Channel is available) Probes: YSI 400 Probes compatible Accuracy: 37.5 ~ 38.0 +/- 0.1°C ( 20~37.4 °C +/- 0.2°C , 38.1~45°C +/-0.2 °C ) Alarm Range: 0°C ~ 45°C Default Setting: Upper Limit 42°C and Lower Limit 35°C |
| Waveforms | SpO2, EtCO2 and EKG |
| Memory | 200 hours |
| Weight | 1.75 kg |
| Dimension (mm) | 200mm (L) x 70mm(W) x 130mm(H) (Without handle) |